Chinese Journal of Medical Instrumentation ; (6): 302-306, 2022.
Artigo em Chinês | WPRIM | ID: wpr-928909

RESUMO

Hyperbaric oxygen therapy is a method of breathing pure oxygen or high-concentration oxygen in a highpressure environment to treat hypoxic diseases and related diseases. According to clinical verification, this therapy has an irreplaceable effect on certain diseases and has gradually become a comprehensive clinical treatment. One of the main methods of certain diseases is widely recognized by the medical field at home and abroad. The development history, treatment principles, key technologies, and future development trends of hyperbaric oxygen are discussed in detail, provide a research direction for the development of hyperbaric oxygen therapy in the future, and at the same time, it has also improved physicians' awareness of hyperbaric oxygen therapy, so as to improving Industry influence.


Assuntos
Oxigenoterapia Hiperbárica , Oxigênio/uso terapêutico , Projetos de Pesquisa
2.
Chinese Journal of Medical Instrumentation ; (6): 278-282, 2022.
Artigo em Chinês | WPRIM | ID: wpr-928904

RESUMO

Mercury sphygmomanometer based on traditional auscultation method is widely used in primary medical institutions in China, but a large amount of blood pressure data can not be directly recorded and applied in scientific research analysis, meanwhile auscultation data is the clinical standard to verify the accuracy of non-invasive electronic sphygmomanometer. Focusing on this, we designed a miniature non-invasive blood pressure measurement and verification system, which can assist doctors to record blood pressure data automatically during the process of auscultation. Through the data playback function,the software of this system can evaluate and verify the blood pressure algorithm of oscillographic method, and then continuously modify the algorithm to improve the measurement accuracy. This study introduces the hardware selection and software design process in detail. The test results show that the system meets the requirements of relevant standards and has a good application prospect.


Assuntos
Auscultação , Pressão Sanguínea/fisiologia , Determinação da Pressão Arterial , Oscilometria , Esfigmomanômetros
3.
Chinese Journal of Medical Instrumentation ; (6): 248-253, 2022.
Artigo em Chinês | WPRIM | ID: wpr-928898

RESUMO

To solve the problem of real-time detection and removal of EEG signal noise in anesthesia depth monitoring, we proposed an adaptive EEG signal noise detection and removal method. This method uses discrete wavelet transform to extract the low-frequency energy and high-frequency energy of a segment of EEG signals, and sets two sets of thresholds for the low-frequency band and high-frequency band of the EEG signal. These two sets of thresholds can be updated adaptively according to the energy situation of the most recent EEG signal. Finally, we judge the level of signal interference according to the range of low-frequency energy and high-frequency energy, and perform corresponding denoising processing. The results show that the method can more accurately detect and remove the noise interference in the EEG signal, and improve the stability of the calculated characteristic parameters.


Assuntos
Algoritmos , Eletroencefalografia , Processamento de Sinais Assistido por Computador , Razão Sinal-Ruído , Análise de Ondaletas
4.
Chinese Journal of Medical Instrumentation ; (6): 455-458, 2008.
Artigo em Chinês | WPRIM | ID: wpr-309550

RESUMO

This article introduces a measuring method for invasive blood pressure based on SecWave technology and a testing method of effectiveness based on IBP database and simulator. Besides, quantified indexes are set out for the accuracy evaluation, such as static pressure accuracy, accuracy of dynamic pressure pulse wave recognition, pulse rate and the response time, so as to provide an important reference method of effectively objective evaluation for invasive blood pressure measurements.


Assuntos
Humanos , Algoritmos , Processamento Eletrônico de Dados , Métodos , Determinação da Pressão Arterial , Métodos
5.
Chinese Journal of Medical Instrumentation ; (6): 439-442, 2007.
Artigo em Chinês | WPRIM | ID: wpr-323237

RESUMO

This article introduces a method of practical noninvasive blood pressure measurement techniques and its key technical points from the view of products , and it gives detailed requirements for us to evaluate the efficiency of these techniques, The evaluation results are presented here by the testing examples guiding us to a correct evaluation of noninvasive blood pressure measurement techniques.


Assuntos
Determinação da Pressão Arterial , Métodos
6.
Chinese Journal of Medical Instrumentation ; (6): 92-95, 2005.
Artigo em Chinês | WPRIM | ID: wpr-241092

RESUMO

The circuit of a monitoring system for respiratory mechanical parameters is designed based on the detection of respiratory flow and pressure. Breaking through the restrictions of traditional methods that can only monitor respiratory rate, this design is able to monitor more than 10 respiratory parameters simultaneously and thus provides a good technical support for improving the properties of homemade monitors.


Assuntos
Humanos , Resistência das Vias Respiratórias , Simulação por Computador , Computadores , Desenho de Equipamento , Monitorização Fisiológica , Métodos , Mecânica Respiratória , Design de Software , Terapia Assistida por Computador , Métodos , Transdutores